PERFORMANCE AND COMPLIANCE
Tests prove it.
Erlab filtration is the most efficient solution for your laboratory.
CONTROL
Afnor NF-X 15-211. The most demanding standard to the world for recirculating fume hoods.
Mandated by AFNOR, the Union de Normalisation Mécanique (UNM), composed of a college of experts in ventilation, public bodies, members of professional unions, has established the standard AFNOR NF X 15-211: 2009. This standard applies to fume hoods (also known as recirculating fume hoods or ETRAFs) designed for all laboratories in which chemical agents subject to an occupational exposure limit value (OELV) are handled.
This standard requires performance criteria related to filtration efficiency, enclosure containment efficiency, face velocity and documentation.
Filtration efficiency.
It refers to the capacity of the filter to retain the harmful molecules handled in the enclosure and guarantees the quality of the air recirculated downstream of the filter.
The effectiveness of enclosure containment.
It defines the capacity of the fume hood to maintain the vapors or particles in the enclosure without them being able to disperse in the laboratory environment.
Download the test results
Filtering fume hoods
Air Velocity.
It refers to the ability of the fume hood to create a dynamic barrier between the manipulator and the handling. For fixed front filtration hoods, the front air velocity at any point of the openings must be between 0.4 and 0.6 m/s. Fume hoods must also be equipped with a continuous ventilation monitoring device.
Documentation: « The Chemical Listing ».
Filtration fume hoods must be accompanied by a booklet containing an exhaustive list of the chemical agents, that the manufacturer certifies, can be handled in the fume hood under the conditions described in standard NF X 15-211: 2009, indicating for each of them :
• The name of the chemical agent, its formula, its CAS number, its boiling point, its molecular weight, its vapor pressure, etc.
• The reference of the adapted filter and its retention capacity in accordance with the NF X 15-211 standard.
• The type of air quality monitoring recommended to control filter efficiency
• The name of the test laboratory that performed the type test.
The result of 50 years of research and development related to filtration technologies, this vademecum is a testament to Erlab’s expertise in molecular and particulate filtration. In compliance with the AFNOR NF X 15-211: 2009 standard, this document, delivered with each fume hood, includes an exhaustive list of the chemical agents, that Erlab certifies, can be handled in the safest possible conditions.
CONTROL
Erlab products meet the most demanding international standards.
Erlab is the inventor and world leader in laboratory chemical filtration products without connection.
Its over 50 years of research and development related to filtration technologies allow Erlab to guarantee perfect safety for all its devices that meet the following international standards:
U.K.:
BS 7989
France:
AFNOR NF X 15-211: 2009
USA:
ANSI/AIHA Z9.5 / ASHRAE 110: 1995
Germany:
DIN 12927